Instrument care, cleaning & sterilization
Correct reprocessing keeps fine cardiovascular and micro instruments precise for years. These general steps follow the structure of ISO 17664; always follow your facility’s validated process and the product IFU.
Point of use
Wipe off blood and tissue immediately with a damp, lint-free cloth. Do not let soil dry on the instrument. Never soak in saline.
Transport
Move instruments to reprocessing in a closed container, kept moist. Protect micro tips with tip guards.
Preparation
Open box locks and ratchets; disassemble multi-part instruments (e.g. removable suction cannulas, sliding-shaft instruments where designed to open).
Manual cleaning
Use a neutral or mildly alkaline enzymatic detergent and soft nylon brushes. Flush lumens and suction tubes with a syringe or water gun. Never use steel wool, wire brushes or abrasive powders.
Ultrasonic cleaning
Clean in a fresh ultrasonic bath per the detergent maker’s instructions. Keep instruments open, do not overload, and do not mix titanium with stainless steel in the same load.
Rinsing
Rinse thoroughly, finishing with purified (deionized or RO) water to avoid spotting and staining.
Drying
Dry completely, including joints and lumens — use filtered compressed air where available.
Inspection
Check under magnification: tip alignment, serrations, TC inserts, cutting edges, box-lock cracks, ratchet function and corrosion. Withdraw damaged instruments for repair.
Lubrication
Apply a water-soluble, steam-permeable instrument lubricant to hinges and box locks. Do not use oils or silicone sprays.
Packaging
Place in trays or pouches with tip guards on fine instruments; do not stack heavy instruments on micro instruments.
Sterilization
Steam sterilize using your facility’s validated cycle (for example pre-vacuum at 134 °C). Instruments must be clean and dry before sterilization.
Storage
Store dry, in a clean area, away from chemicals. Let instruments cool to room temperature before use or storage.
Avoid
- Leaving instruments soiled or soaking for long periods
- Saline, bleach, chlorine or iodine solutions — they cause pitting
- Steel wool, wire brushes or abrasive cleaners
- Using needle holders with sutures heavier than their rating
- Cutting wire, plastic or suture material with fine tissue scissors
- Dropping or throwing instruments into trays
Special notes
Micro & spring-handle instruments
Clean with tips closed in a protective rack or with tip guards on; inspect tips under magnification after every cycle.
Tungsten-carbide jaws and blades
TC inserts are very hard but brittle: avoid impact and avoid closing jaws on hard objects.
Titanium instruments
Clean separately from stainless steel. Colour changes on anodized titanium are cosmetic.
Long-shaft MIS / VATS instruments
Flush the shaft through the flush port where present, and work the mechanism during cleaning.